Measurement is a value and an interval, not a single number

The metrology spine every other Education First course silently depends on: SI units, significant figures, calibration, accuracy versus precision, and how component uncertainties combine into a reported value plus or minus an interval.

Learning progress

Progress is private to this browser.

Foundation

A number with no unit and no interval is not evidence.

Every downstream Education First course, a certificate of analysis row, a blood biomarker reference interval, a reconstituted concentration, rests on the same metrology spine: units are agreements anchored to nature, and a reported value only means something alongside its uncertainty.

Uncertainty budget

How named input uncertainties, each carried by its sensitivity coefficient, combine into one combined standard uncertainty, and how a coverage factor expands it into the interval a result is reported with.

Modeled
0.40a.u.Type A: the standard deviation of the mean of repeated readings. Enters directly, c₁ = 1.
0.30a.u.Type B: from the reference’s calibration certificate. Enters directly, c₂ = 1.
0.40°CType B: how well the temperature is known, in degrees Celsius.
0.50a.u. per °CThe sensitivity coefficient dY/dT: how far the reading moves per degree. It carries u₃ from degrees into the reading’s unit.
Coverage factor k
Multiplies uc into the expanded uncertainty. For a normal distribution k = 1 covers about 68 percent, k = 2 about 95 percent, k = 3 about 99.7 percent.
10.00 ± Ushare of the combined variance, (cᵢuᵢ)² / uc²u₁ repeatability55.2 %u₂ reference calibration31.0 %u₃ temperature13.8 %

Combined uc

0.539a.u.

√Σ(cᵢuᵢ)², root of the sum of squared contributions.

Expanded U

1.077a.u.

k × uc, k = 2.

Reported interval

10.00 ± 1.077a.u.

c₁u₁ repeatability

0.400a.u.

c₁ = 1.00; 55.2 % of the variance.

c₂u₂ reference calibration

0.300a.u.

c₂ = 1.00; 31.0 % of the variance.

c₃u₃ temperature

0.200a.u.

c₃ = 0.50 a.u. per °C; 13.8 % of the variance.

The components share the budget

Each input enters through its sensitivity coefficient: c₃ = 0.50 a.u. per °C turns u₃ = 0.40 °C into 0.200 a.u. Combined in quadrature the three contributions give uc = 0.539 a.u.; added linearly they would give 0.900, so quadrature is the smaller, and correct, combination for independent inputs. With k = 2 the result is reported as 10.00 ± 1.077 a.u., about 95 percent coverage if the combined distribution is normal. Reducing the repeatability (u₁) contribution, which carries 55 percent of the variance, would shrink U the most.

Uncertainty budget, modeled. Contributions in arbitrary units: repeatability 0.400, 55.2 percent of the variance; reference calibration 0.300, 31.0 percent of the variance; temperature 0.200, 13.8 percent of the variance. Combined standard uncertainty 0.539, coverage factor k = 2, expanded uncertainty 1.077. The result is reported as 10.00 plus or minus 1.077 a.u. These are parameters you set, not measured for any lot.

Method and sources

Y = X₁ + X₂ + c₃(T − T₀); uc = √((c₁u₁)² + (c₂u₂)² + (c₃u₃)²); U = k × uc

What it shows: An uncertainty budget: named inputs, each with a standard uncertainty and a sensitivity coefficient, combine into one combined standard uncertainty, and a coverage factor expands it into the interval a result is reported with.

How to read it: Each input’s contribution is its sensitivity coefficient times its standard uncertainty, in the reading’s unit. The bracket on the ticked axis spans the model reading plus or minus U; the bars show each contribution’s share of the combined variance, (cᵢuᵢ)² / uc². The tiles print uc, U, and each contribution.

Assumptions: The three inputs are uncorrelated (GUM 5.1, no covariance terms); repeatability and the reference calibration enter with unit sensitivity coefficients; the coverage percentages assume the combined distribution is normal. The reading of 10 a.u., the uncertainties, and the temperature coefficient are parameters you set, in arbitrary units.

Sources:
A model of the GUM combination rule that you drive. Not measured data for any lot, and not a potency, efficacy, or dosing claim.

A value axis ticked from 2 to 18 arbitrary units with a bracket spanning 10.00 plus or minus the expanded uncertainty, and three bars, one per named contribution, each as long as its share of the combined variance.

Check yourself

Separate the definition from the measurement.

A learner sees that the SI second is defined by an exact physical constant and concludes every clock reading in seconds must therefore be exact.

Choose the next move

Pick an answer to see why.

The foundation spine, ten concepts.

Foundation before combination before reporting: value-and-interval, SI units, derived units, significant figures, trueness versus precision, and calibration and traceability, then Type A, Type B, combination, and the coverage factor.

A measurement is a value plus an interval

A measurement result is an estimate of the measurand together with a stated uncertainty and unit.

Measured or defined
The measurand, the quantity intended to be measured, plus the dispersion of values reasonably attributable to it.
Method
Report the estimate, the expanded uncertainty, the coverage factor, and the unit together.
Units
The unit of the measurand, plus a dimensionless relative uncertainty for the interval.

Limitations

  • An interval characterizes dispersion; it does not certify the value is correct or fit for any purpose.

Pitfalls

  • Treating a displayed number as exact because it has many digits
  • Ignoring the missing unit or interval

Allowed

A measurement is a value, a unit, and an interval.

Forbidden

This is the exact value; accurate to the last digit.

SI base units and their defining constants

The International System of Units, whose seven base units are the second, metre, kilogram, ampere, kelvin, mole, and candela.

Measured or defined
The link between a unit and a fixed natural constant, for example the second from the caesium hyperfine frequency.
Method
Fix the numerical value of the defining constant exactly; the unit follows from that fixed value.
Units
The seven base units and their symbols; defining constants in their SI units.

Limitations

  • A definition sets the unit; it does not remove the uncertainty of any real measurement made in that unit.

Pitfalls

  • Assuming a value in SI units is exact because the unit is exactly defined
  • Confusing the exact constant with a measured quantity

Allowed

The second is defined by the caesium frequency.

Forbidden

SI units make measurements exact.

Derived units and dimensional consistency

A derived unit is a product of powers of base units, for example the newton as kilogram metre per second squared.

Measured or defined
The dimensional structure of a quantity and whether both sides of an equation carry the same dimensions.
Method
Express each quantity in base units, then confirm the dimensions match across the equation.
Units
Derived SI units (newton, pascal, joule, watt) expressed in base units; prefixes for scale.

Limitations

  • Dimensional consistency is necessary but not sufficient; a dimensionally correct equation can still be numerically wrong.

Pitfalls

  • Adding quantities with different dimensions
  • Dropping a prefix and shifting a value by orders of magnitude

Allowed

The newton is kilogram metre per second squared.

Significant figures and rounding

Significant figures are the digits that carry meaning about the value; the digit count is a crude, implicit uncertainty statement.

Measured or defined
The number of justified digits and the last place to which a value should be reported.
Method
Round the uncertainty to one or two significant figures, then round the value to the same last place.
Units
Same unit as the value; the count of significant figures is dimensionless.

Limitations

  • Significant figures approximate the interval; they do not replace an explicit stated uncertainty.

Pitfalls

  • Reporting more digits than the uncertainty justifies (false precision)
  • Rounding the value to fewer places than the uncertainty (lost information)

Allowed

Report the value to the last place of its uncertainty.

Forbidden

Extra digits make it more accurate.

Trueness, precision, and error versus uncertainty

Trueness is closeness of the mean to the reference value (bias); precision is closeness of repeated results to each other (scatter). Error is a single unknown offset; uncertainty is a characterized interval.

Measured or defined
The bias of the mean and the dispersion of repeated observations, treated as separate quantities.
Method
Estimate the mean and its offset from a reference (trueness) and the standard deviation of repeats (precision).
Units
Same unit as the measurand for bias and spread; dimensionless when expressed relative.

Limitations

  • High precision does not imply trueness; a tight cluster can sit far from the reference.

Pitfalls

  • Calling a precise-but-biased instrument accurate
  • Treating an unknown offset as if it were zero

Allowed

Precise does not mean true.

Forbidden

These repeats prove the instrument is accurate.

Calibration and metrological traceability

Calibration relates instrument readings to reference values; traceability is the property that a result can be related to a stated reference through a documented unbroken chain of calibrations, each with an uncertainty.

Measured or defined
The correction and uncertainty each calibration link contributes on the way from a reading back to the SI.
Method
Compare against a reference standard of stated uncertainty; propagate the reference uncertainty into the result.
Units
Same unit as the measured quantity; each link’s contribution as a standard uncertainty.

Limitations

  • Traceability establishes the chain and its uncertainty; it does not by itself prove a specific reading is correct.

Pitfalls

  • Treating an uncalibrated reading as unbiased
  • Assuming a traceable label removes uncertainty rather than quantifying it

Allowed

Each calibration link adds uncertainty.

Forbidden

Traceable means exact.

Type A evaluation of standard uncertainty

A Type A evaluation obtains a standard uncertainty by the statistical analysis of a series of observations.

Measured or defined
The standard deviation of the observations and the standard deviation of their mean.
Method
Estimate the sample standard deviation s; the standard uncertainty of the mean equals s divided by the square root of n; degrees of freedom equal n minus 1.
Units
Same unit as the observed quantity, or dimensionless when relative.

Limitations

  • Type A captures observed scatter only; it does not capture a bias the repeats never sampled.

Pitfalls

  • Using the sample standard deviation instead of the standard deviation of the mean
  • Assuming more repeats fix an unsampled systematic effect

Allowed

Type A uncertainty comes from repeated observations.

Forbidden

More repeats make the value exact.

Type B evaluation and assumed distributions

A Type B evaluation obtains a standard uncertainty by means other than repeated observation, such as a certificate, a resolution, or a handbook bound, using an assumed distribution.

Measured or defined
The standard uncertainty implied by a half-width under an assumed distribution.
Method
Divide the half-width by the distribution divisor: square root of 3 for rectangular, square root of 6 for triangular, square root of 2 for U-shaped, or the bound divided by k for a normal or expanded bound.
Units
Same unit as the bound, or dimensionless when relative.

Limitations

  • The result depends on the assumed distribution, which is a judgement; a wrong distribution mis-sizes the component.

Pitfalls

  • Using the half-width directly as a standard uncertainty (skipping the divisor)
  • Assuming rectangular when the source implies normal

Allowed

A rectangular half-width divides by the square root of three.

Forbidden

The bound is the uncertainty.

Combining component uncertainties

The combined standard uncertainty is the positive square root of the combined variance, obtained by summing squared component contributions for uncorrelated inputs.

Measured or defined
The combined variance of the result and which component dominates it.
Method
For each component, multiply the standard uncertainty by its sensitivity coefficient; the combined variance is the sum of the squared contributions; take the square root.
Units
Same unit as the result, or dimensionless when relative.

Limitations

  • The simple root-sum-square assumes uncorrelated inputs; correlation adds covariance terms and changes the result.

Pitfalls

  • Adding component uncertainties linearly instead of in quadrature
  • Ignoring correlation between shared-source inputs

Allowed

Components combine in quadrature.

Forbidden

The biggest component is the total uncertainty.

Coverage factor and expanded uncertainty

Expanded uncertainty U equals a coverage factor k times the combined standard uncertainty; it defines an interval expected to contain a large fraction of the reasonably attributable values.

Measured or defined
The half-width of the reported interval at a chosen coverage probability.
Method
Multiply the combined standard uncertainty by k; report the value as estimate plus or minus U with k and the coverage probability.
Units
Same unit as the result, or dimensionless when relative; k is dimensionless.

Limitations

  • The coverage probability for a given k assumes an approximately normal result; a heavy-tailed or low-degrees-of-freedom case needs a larger k.

Pitfalls

  • Reporting U without stating k
  • Assuming k equals 2 always gives exactly 95 percent (it gives 95.45 percent under normality, and less when degrees of freedom are low)

Allowed

Report value plus or minus U at k equals 2, about 95 percent coverage.

Forbidden

The true value is inside this interval; this interval proves the product’s potency.

Operate the concepts.

Six secondary instruments, each a concept you drive rather than read. They follow the spine: define the units, separate trueness from precision, report only justified digits, convert a stated bound into a standard uncertainty, tell three different intervals apart, and trace the chain back to the SI.

SI base units, fixed by constant

Which exact defining constants each of the seven SI base units depends on, and why an exact definition does not make a reading in that unit exact.

Exact by definition, from the SI Brochure; the dependencies are derived from its definitions.

second
Select a base unit to ink every defining constant its definition depends on.
defining constant, exact valueSI base unitΔνCs9,192,631,770 Hzc299,792,458 m/sh6.62607015 × 10⁻³⁴ J se1.602176634 × 10⁻¹⁹ Ck1.380649 × 10⁻²³ J/KNA6.02214076 × 10²³ mol⁻¹Kcd683 lm/Wsecondmetrekilogramamperekelvinmolecandelanamed in the definitionenters through the units it uses

Defining constant

ΔνCs

caesium-133 hyperfine transition frequency

Exact value

9,192,631,770 Hz

Exact by definition; no uncertainty.

Depends on

ΔνCs

1 of 7 defining constants.

Unit fixed

second (s)

The second is fixed by the caesium-133 hyperfine transition frequency

The SI Brochure writes it as 1 s = 9 192 631 770 / ΔνCs, and it depends on no other defining constant. The definition fixes the second; a real reading in seconds still carries the uncertainty of the instrument that made it.

The SI unit second is fixed by the caesium-133 hyperfine transition frequency, ΔνCs, exactly 9,192,631,770 Hz, and on no other defining constant. Exact by definition, not a measured value.

Method and sources

1 s = 9 192 631 770 / ΔνCs

What it shows: The 2019 SI fixes seven defining constants exactly and defines each base unit through them. This graph draws, for every base unit, each constant its definition depends on.

How to read it: Constants on the left, units on the right. Selecting a unit inks its edges: solid for the constant its definition names, dashed for the constants that enter through the units that definition is written in. The coefficient is computed from the exact values.

Assumptions: The values are the fixed numerical values of the 2019 SI and carry no uncertainty by construction. Here k is the Boltzmann constant, not the coverage factor k used elsewhere on this page. The candela’s constant is the luminous efficacy of monochromatic radiation of frequency 540 THz.

Sources:
A definition fixes the unit; it does not make any real measurement in that unit exact, and it says nothing about any lot. Not a potency, efficacy, or dosing claim.

A dependency graph: the seven defining constants on the left, the seven SI base units on the right, and an edge from each constant to every unit whose definition depends on it. The selected unit’s edges are inked.

Trueness and precision

Why bias and scatter are independent axes of measurement quality, why both are estimated from the sample, and why error and uncertainty are different objects.

Modeled
0.05ring radiiWhere the process aims, measured from the reference center. The bias tile is estimated from the shots, not read from this.
0.06ring radiiThe shots' standard deviation about their own mean; the scatter tile reads it back from the shots.
Overlay
Error is one arrow from the reference to the sample mean; uncertainty is a band around the result.
0.250.500.751.00ring radii

Bias of the mean

0.06ring radii

Limit 0.25, the dashed ring.

Scatter s

0.06ring radii

Limit 0.12, the dotted ring around the mean.

Verdict

true and precise

From the two values above, each at or within its limit.

Bias 0.06 within its limit, scatter 0.06 within its limit

The sample mean sits within the bias limit and the shots group within the scatter limit: the only quadrant where a single result is also likely to be close to the truth. The arrow is the estimated systematic error, the bias: one offset from the reference to the sample mean.

Trueness and precision target, modeled. Aim offset 0.05 and spread 0.06 ring radii. From the twelve shots, the bias of the mean is 0.06 against a limit of 0.25 and the scatter is 0.06 against a limit of 0.12, so the cluster is true and precise. The overlay shows error. These are parameters you set, not measured for any lot.

Method and sources

x̄ = (1/n) Σ xᵢ; bias = |x̄ − x_ref|; s = √( Σ |xᵢ − x̄|² / (n − 1) )

What it shows: A target for measurement quality. Trueness is bias, how far the mean of the shots sits from the reference. Precision is scatter, how tightly the shots group about their own mean. They are independent: a cluster can be tight yet biased, or centered yet loose.

How to read it: The cross is the reference; the hollow ring marks the sample mean of the twelve shots; the small open diamond is the aim. The dashed ring around the reference is the bias limit and the dotted ring around the mean is the scatter limit. The verdict reads the two printed values against those limits.

Assumptions: Twelve shots from a fixed standard-normal pattern, scaled so their sample standard deviation equals the spread exactly; the pattern’s own mean is kept, so the sample mean differs from the aim by a sampling error that grows with spread. Limits 0.25 and 0.12 ring radii are teaching cuts, not measured limits. The band radius is 2 s.

Sources:
A model of trueness versus precision that you drive. The shots are not any lot’s replicate set. Precise does not mean true. Not measured data, and not a potency, efficacy, or dosing claim.

Four rings at 0.25, 0.5, 0.75 and 1 ring radius around a reference cross, with a dashed bias-limit ring. Twelve dots are the shots; a hollow ring marks their sample mean with a dotted scatter-limit ring around it. An arrow from the reference to the mean shows error, or a dashed band around the mean shows uncertainty.

Significant-figure reporter

How an uncertainty decides which digits of a value are worth reporting, and which are false precision.

Modeled
The unrounded result in units of your choosing: 0 to 9.99999, at most five decimals.
0.0123Logarithmic: equal travel is an equal ratio, one decade from 0.0001 to 0.001 as from 0.01 to 0.1. Same units as the value.
raw value over reported value, aligned by decimal place1.23456rawfalse precision1.235± 0.012reported, justified by uncertaintyreported: 1.235 plus or minus 0.012

Rounded uncertainty

0.012

Two significant figures.

Last justified place

thousandths

10⁻³

Reported result

1.235 ± 0.012

False-precision digits

2

Raw digits bracketed on the strip.

2 trailing digits not justified

The uncertainty 0.0123 rounds to 0.012 at two significant figures, which places the last justified digit in the thousandths. The value 1.23456 therefore reports as 1.235 plus or minus 0.012; the bracketed raw digits are precision the uncertainty does not support.

Significant-figure reporter, modeled. The uncertainty 0.0123 rounds to 0.012 at two significant figures, and the value 1.23456 rounds to the same last place, the thousandths, so the result is reported as 1.235 plus or minus 0.012. These are values you set, not measured for any lot.

Method and sources

u_reported = round(u, 2 sig. figs); x_reported = round(x, place of u_reported)

What it shows: A reporting rule made operable: an uncertainty is rounded to two significant figures, and the value is rounded to the same last decimal place that rounded uncertainty justifies.

How to read it: The strip sets the raw value’s digits, exactly as entered, over the reported value’s digits, aligned by decimal place. The reported row is the result digit for digit; raw digits below the justified place are bracketed as false precision. The tiles print the rounded uncertainty, the last justified place, the reported result and the count of unjustified raw digits.

Assumptions: The uncertainty is rounded to two significant figures (the GUM allows one or two) and the value to the place of that rounded uncertainty, with round half up on the decimal digits. The uncertainty control moves in steps of 0.01 in log10 and holds its value at three significant figures. Values and uncertainties are the reader’s own.

Sources:
A deterministic rounding rule that you drive. Extra digits past the justified place do not make a result more accurate. Not measured data for any lot, and not a potency, efficacy, or dosing claim.

Two rows of digits aligned by decimal place: the raw value as entered, with the digits the uncertainty does not justify bracketed as false precision, and beneath it the reported value, which is the result digit for digit.

Half-width to standard uncertainty

How a quantity known only to lie within plus or minus a becomes a standard uncertainty u, and why the answer depends on the distribution you assume.

Modeled
Rectangular
How you believe the value is spread between the bounds. The normal is offered at each coverage factor k.
0.70relative unitsThe stated bound: the value lies within plus or minus a.
plus or minus u−a+arelative units, self-referential (numberless)

Rule

u = a / √3

Rectangular assumption.

u as a fraction of a

0.577a

Standard uncertainty u

0.404relative units

From a = 0.70.

Rectangular: the bound gives about 0.577 a

Under the rectangular assumption the value is equally likely anywhere between the bounds, so u = a / √3 and u is about 0.577 a. Every curve encloses the same probability, one; the bound a is the same in every case and only the assumed shape changes, which is why the bound is not the uncertainty.

Half-width to standard uncertainty, modeled. Assumed distribution Rectangular. The standard uncertainty is the half-width a divided by the square root of 3, about 0.577 of a. The half-width and the assumed shape are values you set, not measured for any lot.

Method and sources

u = a / √3 | a / √6 | a / √2 | a / k

What it shows: A converter from a stated half-width a (a quantity known only to lie within plus or minus a) into a standard uncertainty u, under a distribution you assume for how belief is spread across that interval.

How to read it: The curve is the assumed probability density, drawn at equal area on one vertical scale: the rectangle at height one half, the triangle peaking at one, the arcsine 1 / (π √(a² − x²)) rising to the bounds, the normal with the bound at k standard deviations. The shaded band marks plus or minus u. The tiles print the rule, u as a fraction of a, and u in the same relative units as a.

Assumptions: Each divisor is the standard deviation of the named distribution with half-width a: rectangular √3, triangular √6, U-shaped (arcsine) √2, normal k when the bound is taken as k standard deviations. The arcsine is drawn up to the frame where it rises without bound, and the normal’s tails are drawn to the ends of the axis. The half-width and the shape are the reader’s own.

Sources:
A model of the half-width conversion that you drive. The divisor is a judgement about the assumed shape, so the bound is not the uncertainty. Not measured data for any lot, and not a potency, efficacy, or dosing claim.

A numberless horizontal axis with the assumed rectangular density drawn at equal area between minus a and plus a, and a shaded band spanning plus or minus u around the center.

Three intervals, one axis

A reference interval, a tolerance interval with its decision rule, and a coverage interval drawn together, and how a guard band turns one measured value into an accept or reject decision.

Modeled
50.0a.u.The single result being judged.
3.0a.u.U = k u with k = 2; the coverage interval is y plus or minus U.
Decision rule
w = U inside the tolerance limits: accept only when the whole coverage interval is inside.
reference interval (population)tolerance interval and decision ruleALAUTLTUcoverage interval, y ± U

Decision

Accept

y is in the acceptance interval.

Coverage interval

47.0 to 53.0a.u.

y ± U, VIM 2.36.

Acceptance interval

43.0 to 57.0a.u.

Guard band w = 3.0 on each side.

Conformance probability

>99.9%

Normal model, u = U / 2.

Accept: y is in the acceptance interval

y = 50.0 with U = 3.0 is in the acceptance interval for guarded acceptance. The probability that the true value conforms is >99.9 percent. The reference interval is drawn on the same axis but plays no part in the decision: it describes a population, not a specification.

Conformity model. Measured value 50.0, expanded uncertainty 3.0, coverage interval 47.0 to 53.0. Guarded acceptance, acceptance interval 43.0 to 57.0. Decision: accept, y is in the acceptance interval. Conformance probability >99.9 percent. Arbitrary units you set, not measured for any lot.

Method and sources

A_L = T_L + w, A_U = T_U − w (guarded acceptance, w = U); p_c = Φ((T_U − y)/u) − Φ((T_L − y)/u)

What it shows: Three objects that all look like intervals: a reference interval describes a population; a tolerance interval with acceptance limits is a decision rule; a coverage interval is one result plus or minus its expanded uncertainty. Drawn on one axis, the decision depends only on the second and third.

How to read it: The top lane is the reference interval (28 to 72). The middle lane is the tolerance interval (T_L = 40, T_U = 60) with the acceptance limits the chosen rule sets and the guard bands between them; the rejection interval is everything outside the acceptance limits. The bottom lane is y plus or minus U. The decision is accept when y is inside the acceptance interval.

Assumptions: The axis is in arbitrary units. The tolerance and reference limits are fixed teaching values; the measured value, U and the decision rule are yours. The conformance probability assumes a normal distribution centered on y with standard uncertainty u = U / 2.

Sources:
A conformity model you drive, in arbitrary units. Nothing here is measured for any lot, and nothing is a potency, efficacy, or dosing claim.

One axis in arbitrary units with three lanes. Top: the reference interval from 28 to 72. Middle: the tolerance limits at 40 and 60, the acceptance limits the decision rule sets, the guard bands between them, and the rejection interval outside. Bottom: the measured value with its coverage interval, y plus or minus U.

Metrological traceability chain

How a reading connects to the SI through a documented chain of calibrations, and how each link’s uncertainty combines into the reading’s by root-sum-square.

Modeled
rung 5 of 5: The instrument reading1 is the SI definition at the top; 5 is the instrument reading at the bottom. Any rung can also be selected on the ladder.
0.20%The relative standard uncertainty the calibration link into the selected rung adds.
+0.02 %+0.05 %+0.10 %+0.20 %SI definition0, exactNational metrologystandard0.020Accredited calibrationlaboratory reference0.054Working standard0.114The instrument reading0.230

Rung

5 of 5

The instrument reading

Cumulative u_c

0.230%

Root-sum-square of the links above.

Expanded U, k = 2

0.460%

Added linearly

0.37%

The wrong combination, for contrast.

The instrument reading: u_c = 0.230 %

The 4 calibration links down to this rung combine in quadrature to 0.230 %; added linearly they would give 0.37 %, which overstates an uncorrelated chain. The link into the instrument reading is the largest and dominates the sum of squares. A traceable label quantifies the uncertainty carried down this chain; it does not remove it.

Traceability ladder, modeled. Selected rung 5 of 5: The instrument reading. Link uncertainty 0.20 percent. Cumulative standard uncertainty 0.230 percent by root-sum-square, 0.37 percent if added linearly. Link values you set, not measured data.

Method and sources

u_c(n) = √(u₁² + u₂² + … + u_n²); U = k × u_c

What it shows: A metrological traceability chain: an unbroken, documented path from the SI definition, through a national metrology standard, an accredited calibration laboratory reference and a working standard, to the instrument reading. Each calibration link adds its own standard uncertainty.

How to read it: Each rung’s bar is drawn to its cumulative standard uncertainty on the percent axis: the root-sum-square of every link from the SI anchor down to it. Select a rung and set the uncertainty of the link into it; every bar below that link redraws.

Assumptions: The links are uncorrelated and enter with unit sensitivity coefficients. The default link values are teaching values you change, not any certificate’s values; a real chain states each link’s uncertainty on its calibration certificate.

Sources:
A model of a traceability chain that you drive; the link values are yours, not any certificate’s. Not measured data for any lot, and not a potency, efficacy, or dosing claim.

A vertical ladder of five rungs from the SI definition at the top to the instrument reading at the bottom. Each calibration link is labeled with the uncertainty it adds, and each rung has a bar drawn to its cumulative standard uncertainty on a percent axis; the SI anchor’s bar has zero length.

Claim boundary

Allowed

  • A measurement is a value, a unit, and an interval.
  • You set the components; the combined and expanded uncertainty are computed from your inputs.
  • The parameters are yours to set, not measured for this lot; this is not a product potency, efficacy, or dosing claim.
  • Report the value as estimate plus or minus U at k equals 2, about 95 percent coverage under a normal approximation.
  • Each calibration link adds uncertainty; traceability quantifies uncertainty, it does not remove it.
  • Precise does not mean true.

Forbidden

  • This interval is the exact value
  • the true value is certainly inside this interval
  • Accurate to within X applied to any catalog item or lot
  • This measurement proves the product’s potency, purity, efficacy, or fitness
  • This is the lot’s measured value for any learner-set or example number
  • Any traffic-light or good/bad scoring of a value or interval
  • Any conformity pass or fail rendered against a real product spec that reads as a verdict on a lot
  • Any diagnostic, dosing, therapeutic, or human-use instruction

Practice bank, 10 prompts

Prompt 1 of 10

0 of 10 reviewed this session

A measurement is a value plus an interval

Identify the three parts of a stated result.

Prompt 1 of 10: A measurement is a value plus an interval.

Source record

Source record

These sources support the metrology vocabulary, formulas, and boundaries taught in this course. They are not cited as evidence that any catalog item changes a biological outcome.

JCGM 100:2008, Evaluation of measurement data, Guide to the expression of uncertainty in measurement (GUM)

BIPM / JCGM

Type A and Type B evaluation, the law of propagation, coverage factor, expanded uncertainty, effective degrees of freedom.

Accessed 2026-07-10

Source

JCGM 200:2012, International vocabulary of metrology (VIM), 3rd edition

BIPM / JCGM

Measurand, measurement uncertainty, trueness, precision, accuracy, metrological traceability definitions.

Accessed 2026-07-10

Source

BIPM SI Brochure, 9th edition (2019)

BIPM

Seven base units, seven defining constants and their exact values, derived units, prefixes.

Accessed 2026-07-10

Source

NIST SP 811 (2008), Guide for the Use of the International System of Units

NIST

Unit style, significant digits, rounding, converted-value rounding.

Accessed 2026-07-10

Source

NIST Technical Note 1297 (1994), Guidelines for Evaluating and Expressing the Uncertainty of NIST Measurement Results

NIST

Combined and expanded uncertainty, coverage factors, practical reporting.

Accessed 2026-07-10

Source

ISO/IEC 17025:2017, General requirements for the competence of testing and calibration laboratories

ISO

Metrological traceability requirement (clause 6.5), calibration and uncertainty in accredited laboratories.

Accessed 2026-07-10

Source

ISO 5725-1, Accuracy (trueness and precision) of measurement methods and results, part 1

ISO

Trueness, precision, repeatability, reproducibility definitions.

Accessed 2026-07-10

Source

ICH Q2(R2), Validation of Analytical Procedures

ICH / FDA

Accuracy, precision (repeatability and intermediate precision), detection limit, quantitation limit.

Accessed 2026-07-10

Source

Educational content only. This course teaches the measurement system, not a substance. It does not provide medical advice, diagnostic evidence, dosing guidance, or evidence that any catalog item changes a biological outcome.

Editorial scope: chemical identity, experimental context, analytical methods, source quality and regulatory boundaries. Source references and stated limitations form part of the material; they should remain attached to quotations or summaries.